Page 1 - Switching DC Power Supply

Model 1900, 1901, 1902 Switching DC Power Supply INSTRUCTION MANUAL

Page 2

8 4 Operating Instructions Safety Precautions ● Never short the remote sensing terminal. ● This power supply is for indoor use only. ● Do n

Page 3 - 1 Safety Summary

9 4.1 Using the Power Supply 4.1.1 Connection To connect the equipment to the power supply, follow the steps below. 1. Check the rating label of

Page 4 - Certification

10 Front Panel Display Test Segment check C.V. indicator check C.C. indicator check Rear control indicator check Return to C.V. Start power su

Page 5

11 Front Panel Display Test Overtemperature protection check Fan check Output off (remote control mode) Table 2 - Self Test Sequence The LED and

Page 6 - Contents

12 4.1.4 Using Both Main and Auxiliary Outputs The power supply has a main output in the rear and an auxiliary output in the front that can be used

Page 7 - 2 Introduction

13 Note: The power supply is factory preset to Normal Mode with maximum current level. 4.2.1 Normal Mode This is the factory preset mode and the po

Page 8 - 3 Controls and Indicators

14 4.2.3 Set Mode First, enter Set Mode by pushing Mode Selection Switch to “Set” position. To define the preset output P1/P2/P3 1. Select the Reca

Page 9 - 3.2 Rear Panel

15 4.2.4 Analog Remote Control Mode Select this mode to control the output voltage and current by Remote Control Connector. Please refer to Section

Page 10 - 4 Operating Instructions

16 4.3.2 Disconnection WARNING: Wrong disconnect sequence will damage the power supply. 1. Disconnect the remote sensing connections. 2. Disconne

Page 11 - 4.1 Using the Power Supply

17 Figure 5 - Pin Numbers (c) Make sure the load is disconnected and the power supply is OFF. (d) Plug the remote connector plug into the analog r

Page 13 - 4.1.3 Control Knobs

18 Method A: Using Two External Variable DC Voltage Sources PIN FUNCTIONS REMARKS 1 Internal DC +5 V Less than 50 mA 2 Voltage Adjust 0 – 5 V

Page 14 - 4.2 Control Modes

19 5. Switch on the power supply. 6. Check the output voltage range of the power supply by varying the external voltage source for voltage adjustme

Page 15 - Section 4.2.3 for details

20 PIN FUNCTIONS REMARKS 1 Internal DC +5 V Resistor end 2 Voltage Adjust Variable part of resistor 3 Current Adjust Variable part of resisto

Page 16 - 4.2.3 Set Mode

21 Shorting Pin 5 to Pin 4 (ground) will turn the output off. When output is off, the CV and CC LED will flash. The current output voltage and curren

Page 17 - 4.3 Remote Sensing

22 5.2 OTP: Overtemperature Protection There is a thermo sensor inside the unit to monitor and prevent the unit from getting too hot. When OTP is t

Page 18 - 4.4 Analog Remote Control

23 Figure 10 - Overload Protection To reset this warning, switch off the unit and remove all connected devices. Switch the unit back on again and do

Page 19 - Figure 6)

24 6 Specifications Models 1900 1901 1902 Output Variable Output Voltage 1 – 16 V 1 – 32 V 1 – 60 V Variable Output Current 0 – 60 A 0 – 30 A 0

Page 20

25 Models 1900 1901 1902 Switching Frequency 65 – 85 kHz 75 – 95 kHz 65 – 85 kHz Tracking Overvoltage Protections O/P 1-5 V: set voltage +2 V O

Page 21 - Figure 7

26 7 Certification CE Compliant CE Declaration of Conformity The power supply meets the requirements of 2006/95/EC Low Voltage Directive and

Page 22

27 8 Service Information Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain an RMA #. R

Page 23

1 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of ope

Page 24

28 9 Limited Two-Year Warranty B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof will

Page 25 - 5.4 Fuse Replacement

(Page intentionally left blank)

Page 26 - 6 Specifications

Printed in Hong Kong v060811 22820 Savi Ranch Pa

Page 27

2 WARNINGS AND CAUTIONS WARNING and CAUTION statements, such as the following examples, denote a hazard and appear throughout this manual. Follow al

Page 28

3 Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with sep

Page 29 - 8 Service Information

4 Contents 1 Safety Summary ... 1 2 Introduction ...

Page 30

5 2 Introduction B&K Precision models 1900, 1901, and 1902 are laboratory grade switching mode DC power supplies with high current output in a

Page 31

6 3 Controls and Indicators 3.1 Front Panel Figure 1 - Front Panel (1) LED panel meter display with CV/CC indicator (2) Rear Control Indicator

Page 32

7 3.2 Rear Panel Figure 2 - Rear Panel (7) Main Output Terminal (max 60 A for 1900 / 30 A for 1901 / 15 A for 1902) Note: Please see Section