Performance Verification Procedure
239
The only parameter that can be user calibrated is the DC gain
accuracy. If any of the other parameters, bandwidth, trigger
sensitivity, or time scale accuracy does not meet published
specifications, the unit must be returned to B&K Precision for repair.
The oscilloscope under test must be warmed up for at least 30
minutes prior to the start of any performance test.
DC Gain Accuracy
Equipment needed
: Fluke calibrator (preferred) or DC power supply,
DMM, splitter, 2 BNC cables
1. Disconnect all cables from the oscilloscope channel inputs.
2. Press the Acquire front panel key.
3. In Acquire menu, press the Mode soft key until Averages
appears.
4. Turn the entry knob until “256” appears.
5. In Acquire menu, press Sampling soft key and set to “Real
Time”.
6. Set CH1 probe attenuation to 1X in CH1 menu.
7. Press Measure button and select Voltage, then go to page
2/4 and select Average.