Menu Operation
84
Fast-Calibration
Fast calibration is ideally used to calibrate the instrument to
remedy the effects of temperature drift causing an offset drift.
It calibrates the center position of each Volt/Div setting, but
not for the full range. This is different compare to
self-calibration, in which the channel scale, offset, and trigger
are calibrated.
I/O Setup
Press UTILITY I/O Setup to display the I/O SETUP menu.
Softkey
Options
Description
Type
USB
Device
Select USB
interface
RS232C
Select RS232C
interface
LAN
Select LAN
interface
Baud
Rate
Available baud
rate: 2400, 4800,
9600, 19200,
38400
Network
Settings
---- Select LAN menu
----
Return to the
UTILITY menu